Standards for UV Disinfection and Sterilisation Dose Requirements
Ultraviolet Disinfection and Sterilisation Dose Requirements
· Disinfecting the surface of objects
– To kill general bacterial vegetative cells: the irradiation dose should reach 10,000 μW·s/cm²
– Killing bacterial spores: should reach 100,000 μW·s/cm²
– Virus inactivation: intermediate between bacterial vegetative cells and spores
– Killing fungal spores: requires a higher dose than bacterial spores; in some cases, an irradiance of up to 600,000 μW·s/cm² is required (generally, pathogenic fungi are less resistant to ultraviolet light than bacterial spores)
– Killing of unspecified target microorganisms: shall not be less than 100,000 μW.s/cm² (Source: ‘Technical Specifications for Disinfection’)
– Killing Escherichia coli: should reach 20,000 μW·s/cm²
– Inactivation of Bacillus subtilis black variant spores: 100,000 μW.s/cm² should be achieved
(GB 15981-1995 Methods and Standards for the Evaluation of Disinfection and Sterilisation Efficacy)
· Indoor air disinfection
– Direct illumination method
· For use indoors without personnel present; suspended or mobile
· The number of UV disinfection lamps installed indoors (30 W UV lamps, with an intensity of >70 μW/cm² at a distance of 1.0 m) shall be no less than 1.5 W per m³ on average, and the irradiation time shall be no less than 30 minutes.
– Indirect illumination method
· High-intensity UV air disinfectors are the preferred choice
· Generally, a 30-minute disinfection cycle upon start-up is sufficient to meet disinfection standards
(Source: ‘Technical Specifications for Disinfection’)
· Disinfection of water and other liquids
– In-water irradiation or out-of-water irradiation
– The water layer should be less than 2 cm thick
– Determine the flow velocity based on the intensity of the UV light source