Specification Requirements for Ultraviolet Disinfection and Sterilisation Products
Specification Requirements for Ultraviolet Disinfection and Sterilisation Products
1.Ultraviolet Disinfection (Sterilisation) Lamps – Low-Pressure Mercury Lamps
– Ultraviolet radiation efficiency: The efficiency with which electrical input power is converted into effective ultraviolet disinfection radiation (%)
– Ultraviolet irradiance: effective ultraviolet germicidal radiation per unit area (μW/cm²)
– UV radiation flux retention rate and lifetime: the attenuation of effective UV disinfection radiation over time
– Ozone output rate: the ratio of the amount of ozone produced per unit of time to the lamp power (g/kW·h)
Radiant irradiance (μW/cm²)
Dose of germicidal disinfectant: μJ/cm² = μW·s/cm²
– Ultraviolet radiation efficiency:Conversion efficiency of input electrical power to effective UV disinfection radiation (%)
(Low-voltage mercury lamps up to 65W range 10% to 23%)
Source: GB/T 19258-2012 Ultraviolet Germicidal Lamps
– Ultraviolet radiation irradiance:Effective UV disinfection radiation per unit area (μW/cm²) (measured at a distance of 1 m)
Rated values for single-ended UV lamps: 24W: 94 μW/cm²; 36W: 147 μW/cm²; 55W: 170 μW/cm²; (Source: GB 28235-2011 Safety and Hygiene Standards for Ultraviolet Air Disinfectors)
Standard straight-tube low-pressure mercury UV disinfection lamps: >30 W: ≥90 μW/cm²; >20 W: ≥60 μW/cm²; 15 W: ≥20 μW/ (Source: ‘Technical Specifications for Disinfection’)
High-intensity UV disinfection lamps: 30W: >170 μW/cm²; 11W: >40 μW/cm²
Standard straight-tube, low-pressure mercury UV disinfection lamp (30W): ≥70 μW/cm²
High-intensity ultraviolet disinfection lamp: ≥200 μW/cm² (Source: GB 15981-1995 Methods and Standards for the Evaluation of Disinfection and Sterilisation Efficacy)
– UV radiation flux retention rate and service life:Decay of effective UV disinfection radiation over time
·2000 h retention rate ≥ 85 %
· Average service life of not less than 5,000 hours; retention rate at end of service life ≥ 65 % (Source: GB/T 19258-2012 Ultraviolet Germicidal Lamps)
· The irradiance of an ultraviolet lamp (30 W) in use should be ≥ 70 μW/cm² (Source: GB 15982-2012 Hygiene Standards for Hospital Disinfection)
· The time taken for the irradiance of an ultraviolet lamp (≥30W) to decrease by 70 μW/cm² is ≥1000 hours
· The time taken for the irradiance value of an ultraviolet lamp (<30W) to decrease by 70% from its initial value is ≥1000 hours (Source: ‘Technical Specifications for Disinfection’)
– Ozone production rate:The ratio of the amount of ozone produced per unit of time to the lamp power (g/kW·h)
· The initial ozone production rate of ozone-free lamps shall not exceed 50 mg/kW·h (Source: GB/T 19258-2012 Ultraviolet Germicidal Lamps)
· Ozone output < 1 mg/h (Source: ‘Technical Specifications for Disinfection’)
· The concentration of ozone generated by low-ozone UV lamps should not exceed 0.3 mg/m³
· The ozone concentration in the working environment of ozone-generating disinfection equipment must be < 0.16 mg/m³ (Source: GB 15982-2012 Hygiene Standards for Hospital Disinfection)
· Other: The maximum permissible 1-hour average ozone concentration is 0.1 mg/m³ (Source: GB/T 18202-2000 Hygiene Standards for Ozone in Indoor Air)
2.Ultraviolet Air Disinfectors – Cabinet-style / Wall-mounted / Portable – Low-ozone UV lamps
– Specification requirements for ultraviolet disinfection (sterilisation) lamps
– Ultraviolet leakage: when measured at a distance of 30 cm, the leakage must not exceed 5 μW/cm²
– Ozone leakage: The maximum permissible average ozone concentration over 1 hour is 0.1 mg/m³
– Air circulation volume: not less than 10 times the applicable volume
– Applicable volume: not less than 30 m³
(Source: GB 28235-2011 Safety and Hygiene Standards for Ultraviolet Air Disinfectors)
For the measurement of ultraviolet energy levels and irradiance, specialist equipment may be usedUltraviolet detection instrument。