Evaluation of the Microbial Killing Efficacy of Ultraviolet Disinfection and Sterilisation Products
– Microbial inactivation tests: Staphylococcus aureus (ATCC 6538), Escherichia coli (8099), Pseudomonas aeruginosa (ATCC 15442), Bacillus subtilis var. niger spores (ATCC 9372), Staphylococcus albus (8032, used for air disinfection), and other bacteria, their spores, and fungal suspensions (Source: ‘Technical Specifications for Disinfection’)
– Indicator organisms for surface disinfection: Escherichia coli (8099), Bacillus subtilis var. niger spores (ATCC 9372) (Source: GB 15981-1995 Methods and Standards for the Evaluation of Disinfection and Sterilisation Efficacy)
– Air disinfection efficacy test: Disinfection time not exceeding 3 hours for a kill rate of ≥99.90% against Staphylococcus albus (strain 8032) or a mortality rate of ≥90% against natural bacteria
(Source: GB 28235-2011 Safety and Hygiene Standards for Ultraviolet Air Disinfectors. Note: The ‘Technical Specifications for Disinfection’ state that disinfection is generally achieved within 30 minutes of commencement.)
– Microbial inactivation efficacy test (maximum load): Staphylococcus aureus (ATCC 6538), Escherichia coli (8099), Pseudomonas aeruginosa (ATCC 15442), Bacillus subtilis var. niger (ATCC 9372) spores, Candida albicans (ATCC 10231), Mycobacterium chelonae subsp. abscessus (ATCC 93326) and other suspensions, as well as poliovirus suspensions
(Source: ‘Technical Specifications for Disinfection’)